Lookup tables

The RMS-MLOG-T10-868/915 is a data logger for use with external NTC's.

Rotronic currently offers the following NTC's:

T10-0001 T10-001 -200...-90 °C 2 m Cryotechnology
T10-0002 T10-002/6 -80...200 °C 2 m Freezers, dry ice, etc.
T10-0003 T10-003/4 -50...200 °C 2 m Standard
T10-0013 T10-003/4 -50...200 °C 2 m Climatic chambers
T10-0113 T10-003/4 -50...200 °C 2 m Climatic chambers with condensation
T10-0004 T10-003/4 -50...200 °C 2 m Pipe monitoring
T10-0005 T10-005 -90...50 °C 5 m Freezers, dry ice, etc.
T10-0006 T10-002/6 -80...200 °C 4 m Standard

The RMS-MLOG-T10-868/915 currently offers the possibility to match the NTC with a lookup table, when using the lookup table, the measurement accuracy is valid throughout the temperature range.

The lookup table is stored on the RMS-MLOG-T10-868/915's memory. This memory is programmed at Rotronic AG. It is not possible to update the lookup tables without returning the device to Rotronic AG. A device will only show the lookup tables that are stored on the device memory independent of the firmware version.

If the lookup table for an NTC is not available, then the user must either update the device (return it to Rotronic AG) or add the R25 and B constant based upon the table below:

R25 & B: -200...-165 °C 54 & 666
R25 & B: -185...-145 °C 26 & 776
R25 & B: -145...-95 °C 8 & 1045
R25 & B: -200...-120 °C 21 & 787
R25 & B: -80...-40 °C 13000 & 3500 13000 & 3500
R25 & B: -40...-30 °C 9950 & 3800 9950 & 3800
R25 & B: -10...150 °C 10100 & 4030 10100 & 4030
R25 & B: -50...-10 °C 11100 & 3610 11100 & 3610
R25 & B: -10....50 °C 9800 & 3890 9800 & 3890
R25 & B: 50...120 °C 10300 & 4065 10300 & 4065
R25 & B: -90...-40 °C 1500 & 2900
R25 & B: -60...0 °C 1100 & 3170
R25 & B: -20...50 °C 100 & 3400

Should the T10 probe be used with the RMS-MLOG-T10-868/915 outside of the temperature range, then the RMS-MLOG-T10-868/915 will show "NaN" in RMS-CONFIG and "---" in the RMS software.

Important:
Please be aware that the RMS-MLOG-T10-868/915 units will only show results when working at  the T10 NTC's operating temperature defined in the table above.

Please find the overview of the lookup tables saved on the RMS-MLOG-T10-868/915 below:

V1.1 R25/B constant All
V1.2/V1.3 R25/B constant
Internal
T10-003/4/5
(Please be aware that the 5 is incorrect, the R25/B constants must be used for the T10-0005)
T10-002 (-80 °C)
T10-001 (-200 °C)
From V1.2
V1.2/V1.3 R25/B constant
Internal
T10-003/4
T10-002 (-80 °C)
T10-001 (-200 °C)
From V2
V1.4/V1.5 R25/B constant
Internal
T10-003/4
T10-002/6
T10-001
T10-005
From V1.3.1
Important:
The lookup tables are stored on the EEPROM of the device and not within the firmware. If the EEPROM needs updating to include new lookup tables, the MLOG-T10 must be returned to Rotronic AG.