Accuracy¶
General¶
The accuracy of a RMS-LOG-W-T10-D is effected by
- Analog-Digital-Convertion (ADC),
- Used sensor type
- Resistance-Temperature-Calculation (based on sensor table or parameters)
- Long term drift of the sensor
- and other influences as e. g. adjustment and its related reference accuracy
Accuracy of the electronic (ADC)¶
| Accuracy ADC | <0.15°C (range of -70...+165°C for T10-0002) |
|---|---|
| Ambient temperature influence | <0.1°C (range of -30...+50°C ambient temperature of the logger) |
Accuracy of the NTC sensor¶
Please see chapter Temperature NTC
Measurement uncertainty¶
To calculate the a measurement uncertainty for an individual application, all known accuracies/failure influences might be known and calculated by Pythagorean Addition of the known accuracies/failure influences.