Carry out an Adjustment

Within RMS-CONFIG it is important to understand how the adjustment function works.

  • A 1 point adjustment is an offset.
  • A 2 point adjustment is an offset and a slope.
  • A >2 point adjustment is an offset, a slope and smoothing.

When adjusting temperature, a maximum of 2 points can be adjusted.

Adjustment process.

  1. Delete all points acquired within the device.
  2. After the defined time for the calibration, add a reference value and click on acquire.
  3. Should a 1 point adjustment be required, press on adjust.
  4. Should a 2 point adjustment be required, after the defined time for the calibration, add a reference value and click on acquire.
  5. Should a 2 point adjustment be required, press on adjust.
  6. Should a >2 point adjustment be required, after the defined time for the calibration, add a reference value and click on acquire. Repeat this step for the desired amount of adjustments.
  7. When all points have been acquired, press on adjust.