Technical Specification

Sensor 1 or 2 RTD sensors (PT100) connectable
Maximum measurement range ±0.2 °C (-200 °C bis 850 °C) without Pt100
Reproducibility measurement channel +/- 0.1°C (measurement range -100 °C to +200 °C of the channel) with the logger at 23 °C ambient temperature
NOTE:
The reproducibility of the PT100 sensor is to be added still.
Reproducibility measurement channel
for maximum range
+/- 0.2 °C (measurement range -200 °C to +850 °C of the channel) with the logger at 23 °C ambient temperature
NOTE:
The reproducibility of the PT100 sensor is to be added still.
Frequency 868 / 915 MHz
RF Range logger to gateway 868 MHz: 20 - 50 m (within a building)
915 MHz: 15 - 25 m (within a building)
Maximum number of logger per gateway (per channel) 60 loggers
Power supply 24 VDC ±10 %
Current consumption < 100 mA
External power supply advised 24 VDC ±10 %, 4 W nominal, > 15 W Limited Power Source
Operation time of the battery The batteries of RMS devices typically last 3 years under the optimal conditions:
- Temperature = 20°C±5°C.
- Wireless signal strength >-60 dBm.
- Limited “read out data” events in the audit trail.
Log memory 44000 of measurement for each channel
Startup time logger < 10 s
Measurement interval >= 10 s
Adjustment 1 or 2 point per channel
Sensor quality - Detection of short circuit on the sensor connectors
- Detection if no sensor connected
Temperature simulator function for validation Fix values for each channel
Operational temperature logger -40 ... 70 °C
Storage temperature -40 ... 30 °C
NOTE:
Small storage temperature range is based on the battery life
Housing Material PC-ABS
IP protection IP65
Weight 240 g