RMS-LOG-L-D¶
The RMS-LOG-L is a RMS logger with the following main functions:
- Request the measurement data from the connected digital probe (1 or 2 measurement values).
- Transmit the measurement data, including time stamp and device information to the RMS Webservice via Ethernet.
-
Store measurement data, including time stamp during an interruption of communication and supply the data afterwards to the RMS.
-
Store the warning and alarm limits, including the hysteresis, defined by the RMS Webservice (the alarm limits are not visible on the RMS-LOG-L-D).
-
The RMS-LOG-L-D will not take any delays programmed in RMS into consideration. This is per design, to ensure that the users near the device see the red screen and hear the buzzer.
-
Notify the user when a warning, alarm, or error occurs with the built in buzzer and colored display.
| Important: The RMS-LOG-L-D cannot be used as a stand alone device without RMS. The RMS-LOG-L-D cannot recognize when the RMS-HCD is setup with dew/frost point and temperature and will show an error. The RMS-LOG-L-D will ONLY store the following configuration from RMS: - High warning, low warning and hysteresis. - High alarm, low alarm and hysteresis. The RMS-LOG-L-D will not recognise warning and alarm delays. As soon as warning/alarm limits are reached the RMS-LOG-L-D will notify the user as configured (red display and buzzer). The values on the display will only show, this cannot be configured: 1. Hundreds: 1xx.xx (if no hundreds, then no value will be shown). 2. Tens: x1x.xx. 3. Ones: xx1.xx. 4. Tenths: xxx.1x. 5. Hundredths: xxx.x1. |
|---|