Technical specifications

The CCA-S-20X is a carbon dioxide probe designed for incubator monitoring applications.

General information

Functions Description
Measured parameters The CCA measures carbon dioxide
Technology NIBR
Range 0...20%CO2
Accuracy 0...8%CO2 ±0.15%CO2
0...20%CO2 ±0.2%CO2
Long term stability ±0.24%CO2/Year
Temperature dependence ±10% of reading
Pressure dependence ±0.15% of reading/hPa
Range of application -20...50°C
0...95 %rh non condensing
700...1200hPa
Startup time 60s
Power supply CCA-S-20X: 3...5 VDC
CCA-S-20X-SET: 12 VDC, 0.5 A
Output signal CCA-S-20X: 0...2.8 V
CCA-S-20X-SET: 4...20 mA
Enclosure material Polycarbonate (housing)
Stainless steel DIN 1.4305 (nut)
Fire protection class Corresponds to UL94-HB
Dimensions CCA-S-20X: 32 mm x 87 mm
Converter box: 100 mm x 77 mm x 40 mm (LxWxH)
IP protection class IP40
Weight CCA-S-20X: 55 g
CCA-S-20X-SET: 255 g

Additional functions with the CCA-S-20X-SET only (via the converter box)

Functions Factory default
Sensor zero Factory adjusted
Sensor span Factory adjusted
Sensor calibration Factory calibrated
4...20 mA loop calibration Factory calibrated

Example Accuracy Overview

Reading: 5.23%CO2.

Temperature: 37°C.

Pressure: 1050 hPa.

Calibration pressure: 1000hPa.

Sensor age: new.

**Measurement Accuracy: **

5.23%CO2 ±0.15%CO2 = 5.08%CO2<x<5.38%CO2.

**Temperature Dependence: **

5.23%CO2 ±10% of reading = 4.707%CO2<x<5.753%CO2.

**Pressure Dependence: **

**Step 1: 5.23%CO2 ±0.15% of reading/hPa = 0.007845%CO2/hPa. **

Step 2: Pressure difference: 1050hPa-1000hPa=50hPa.

Step 3: 0.007845*50=0.39225%CO2.

Total: 4.83775%CO2<x<5.62225%CO2