Calibration Option 2

This option is valid for any company wishing to calibrate an RMS wireless logger.

Important:
The wireless logger needs to be deactivated within the RMS software.
With RMS-Config, only 1 wireless device can be added at a time.

Requires:

1 * RMS wireless logger

1 * RMS-Config software.

1 * RMS-DONGLE-868/915.

Step 1 Place the wireless logger in the calibration environment.
Plug the RMS-DONGLE-868/915 a computer USB port.
Open the RMS-CONFIG software.
Step 2 Open the RMS-CONFIG software. Click on "Device" > "Search" > "Wireless devices":
calibration option 2 - 1
Step 3 Unpair the wireless logger: push on the button for 8s. The wireless logger will blink 3 x red.
Push on the button once: the logger will blink orange and connect to RMS-CONFIG:
calibration option 2 - 2
Step 4 Click on the Adjust tab:
calibrate digital probe with RMS-CONFIG3
Step 5 When possible, select the measurement value to be calibrated:
calibrate digital probe with RMS-CONFIG4
Step 6 Follow the internal calibration procedures.
Enter the reference value shown on the reference and click on acquire:
calibrate digital probe with RMS-CONFIG5
Step 7 The acquired points will be added to acquired points:
calibrate digital probe with RMS-CONFIG6
Step 8 Carry out all of the necessary calibration points.
The data found within the acquired points is the calibration data with the as found values.
Important:
This is only a calibration and not an adjustment.

Adjust: Click on adjust to adjust the probe. Why Adjust?

Delete point: Delete an acquired point if required.

Delete all points: Delete all points if required.

Restore factory adjust: Restore the factory adjustment if required.