Calibration Option 2¶
This option is valid for any company wishing to calibrate an RMS wireless logger.
| Important: The wireless logger needs to be deactivated within the RMS software. With RMS-Config, only 1 wireless device can be added at a time. |
|---|
Requires:
1 * RMS wireless logger
1 * RMS-Config software.
1 * RMS-DONGLE-868/915.
| Step 1 | Place the wireless logger in the calibration environment. Plug the RMS-DONGLE-868/915 a computer USB port. Open the RMS-CONFIG software. |
|---|---|
| Step 2 | Open the RMS-CONFIG software. Click on "Device" > "Search" > "Wireless devices": ![]() |
| Step 3 | Unpair the wireless logger: push on the button for 8s. The wireless logger will blink 3 x red. Push on the button once: the logger will blink orange and connect to RMS-CONFIG: ![]() |
| Step 4 | Click on the Adjust tab: ![]() |
| Step 5 | When possible, select the measurement value to be calibrated: ![]() |
| Step 6 | Follow the internal calibration procedures. Enter the reference value shown on the reference and click on acquire: ![]() |
| Step 7 | The acquired points will be added to acquired points: ![]() |
| Step 8 | Carry out all of the necessary calibration points. The data found within the acquired points is the calibration data with the as found values. |
| Important: This is only a calibration and not an adjustment. |
|---|
Adjust: Click on adjust to adjust the probe. Why Adjust?
Delete point: Delete an acquired point if required.
Delete all points: Delete all points if required.
Restore factory adjust: Restore the factory adjustment if required.