General information

Functions Description
Measured parameters The CCD measures carbon dioxide and has an integrated pressure sensor for pressure compensation.
Range of application 0…50 °C, 0…95 %rh non-condensing
Accuracy ±50 ppm ± 3 % of read value @ 0-2000 ppm
±10 % of read value @ 2000-10'000 ppm
Ambient pressure compensation 300-1100 hPa
Diameter 32 mm (1.26")
Length 76 mm (3")
Housing material Polycarbonate, black
Standard filter Stainless steel
Sensor NDIR
Weight 55 g (0.35 oz)
Power Supply The clip is created to be used directly connected to and powered by a RMS LAN Logger. Alternative a AC3001 (3.3 V or 5.5 V) can be used.
The RMS Clip CO2 need to be powered permanently because the measurement principle. Therefore, it is not recommended to use a RMS RF Logger. Its battery capacity would drop down drastically from years to few days of permanent use.
The heating up phase is app.300 sec. The measurement interval is 16 sec. or higher. Lower intervals are set automatically to 16 sec.
Digital interface The CCD probe has a UART interface, via which the probe can be addressed directly using the MODBUS protocol. To connect the probe to a PC, an AC3001 service cable is required.
Using the ROTRONIC RMS-CONFIG software, the following functions are available via the UART interface.
Cable length maximum 5 m

Factory defaults

Configurable settings Factory default
Unit of measurement (metric/English) Metric
Psychrometric calculation None
Communication protocol MODBUS
MODBUS address 0
Device name CCD
Functions Factory default
CO2 adjustment Factory adjusted
CO2 ABC Enabled, interval is set to 8 days
CO2 measurement range 0…2’000 ppm, scalable 0…5’000/10’000 ppm
CO2 measurement period 16 s
CO2 filter Filter and APC enabled
CO2 simulator Disabled