The probe can be addressed directly using the MODBUS protocol, to request the measured CO2 values. The parameters highlighted in yellow should be selected according to the table below.

Important:
In order to communicate directly with the digital probe via MODBUS, please use the E2-05XX-MOD cable. The digital MODBUS protocol is delivered over RS485.

Command: (MSB first)

Probe address Command ID Start address Number of records MODBUS CRC1
0x00 0x04 0x00 0x00 0x00 0x00 0x0000

1 CRC calculation according to MODBUS standard

Probe address: 0…255, if '0' then all probes respond regardless of the address.

0x00 0x04 Serial number CO2
0x00000000 0x0000
0x00 0x02 Serial number
0x00000000

Response

0x00 0x04 0x00 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00 0x0000

Data format

Serial number (unsigned int): 0…232-1

Response in the event of an error

0x00 0x84 0x00 0x0000
0x02 Start address and number of records invalid
0x03 Number of records invalid

Sensor faults

In the event of a short circuit or open circuit at the probe, a value of 19999 is output for the measured value concerned.

Coils Bit Read/Write 1 … 9’999 0x01 Read Coils
0x05 Write Single Coil
0x0F Write Multiple Coils
Discrete inputs Bit Read Only 10’001 … 29’999 0x02 Read Discrete Inputs
Input registers 16-bit Read Only 30’001 … 39’999 0x04 Read Input Register
Holding registers 16-bit Read/Write 40’001 … 49’999 0x03 Read Holding Registers
0x06 Write Single Register
0x10 Write Multiple Registers
Note:
Coils and registers in Modbus are addressed starting at zero. Therefore coils numbered 1…16 are addressed as 0…15 or registers numbered (e.g. 10'001…10'016 are addressed as 10'000…10'015)

Connection timeout: When TCP connection is open and for more than 30 seconds no communication is performed, the device automatically closes the TCP connection. In case of a communication interruption, this prevents the socket being blocked.