Skip to content

Technical specification

Device type Carbon dioxide probe
Range of application 0…+50 °C / 0…95 %rh (non-condensing)
IP protection IP40
Power supply / connection
Power supply (VDD) 3.3…5.5 VDC ±0 %
Current consumption (16 s interval) 20 mA (average). 260 mA (peak)
Polarity protection Mechanical protection
Battery life 2.7 days with a 10 s or 60 s measurement interval
CO2 measurement
Sensor Infrared (NDIR)
Parameter CO2 concentration (ppm/%)
Measurement range 0…2’000/5’000/10’000 ppm
Resolution 1 ppm
Measurement accuracy (after minimum 3 weeks ABC) ±50 ppm ±3 % of read value @ 0…2’000 ppm ±10 % of read value @ 2000…10,000 ppm
Medium Air and non-aggressive gases
Ambient pressure and temperature compensation Automatic (300…1100 hPa)
Start time / measurement interval
Startup time <300 s
Measurement interval 16 s
Response time T63 130 s when descending
87 s when ascending
Digital interface
Type UART
Protocol MODBUS RTU
Factory default Baud rate 19200, tolerance 2 %
Parity: none
Data bits: 8
Stop bits: 1
Flow control: none
Logic levels Logic 0: ≤0.4 V
Logic 1: ≥2…2.5 V
Maximum cable length 5 m
General specifications
Housing material Polycarbonate (housing)
Thumb screw material Stainless steel, DIN 1.4305
Weight 55 g
Conformity
CE / EMC EMC Directive: 2014/30/EU
EN 61000-6-1:2007
EN 61000-6-2:2005
EN 61000-6-3:2007+A1:2011+AC:2012
EN 61000-6-4:2007+A1:2011
EN 61326-1: 2013
Performance criterion: www.rotronic.com
Fire protection class Corresponds to UL94-V2
Soldering Lead free (RoHS Directive 2011/65/EU)
FDA/GAMP directives Compatible
Accepted environmental conditions
Storage/transportation 0…50 °C / 0…95 %rh, non-condensing
Range of application (electronics) 0…50 °C / 0…95 %rh, non-condensing