Exchange Digital Probe¶
The logger is compatible with the following digital probes:
When exchanging a digital probe from one type to another, the logger will show a sensor error. Follow these steps to ensure that the logger reads the digital probe correctly:
- Archive the existing measuring point(s).
- Remove the digital probe.
- Reset the logger (push on the button for a minimum of 8s).
- Remove all power suppliers (PoE, 24V and batteries).
- Wait for 5s.
- Add the required power supplies (PoE, 24V and batteries).
- Add the new digital probe.
- Add the new measuring point(s).
The sensor error should have disappeared and the new measurement value should be visible within RMS.