Skip to content

Customer adjustment

General

The customer can compensate possible deviations of the temperature measurement by an own adjustment. The customer uses temperature values as reference points for his adjustment.

1-point adjustment

1-point adjustment corresponds to a offset over the whole measurement range.

Calculation
Offset temperature TOffset = TRef - TAct
TRef = reference temperature
TAct = actual measured temperature (live from the device or noted by the customer)
End temperature TEnd = T(R) + TOffset
T(R) = temperature (already corrected)

2-point adjustment

For the 2-point adjustment, the customer can acquire two points.

Calculation
Slope Slope
Offset intercept
TRef1 = reference temperature 1
TRef2 = reference temperature 2
TAct1 = measured temperature 1 (live from the device or noted by the customer)
TAct2 = measured temperature 2 (live from the device or noted by the customer)
End temperature TEnd = m * T(R) + n
T(R) = temperature (already corrected)