Technical Specification¶
| Sensor | 1 or 2 RTD sensors (PT100) connectable |
|---|---|
| Maximum measurement range | ±0.2 °C (-200 °C bis 850 °C) without Pt100 |
| Reproducibility measurement channel | +/- 0.1 °C (measurement range -100 °C to +200 °C of the channel) with the logger at 23 °C ambient temperature NOTE: The reproducibility of the PT100 sensor is to be added still. |
| Reproducibility measurement channel for maximum range |
+/- 0.2 °C (measurement range -200 °C to +850 °C of the channel) with the logger at 23 °C ambient temperature NOTE: The reproducibility of the PT100 sensor is to be added still. |
| Interface | 10Base-T / 100Base-TX Ethernet |
| Power supply | 24 VDC ±10 % |
| Current consumption | < 100 mA |
| External power supply advised | 24 VDC ±10 %, 4 W nominal, < 15 W Limited Power Source |
| PoE | 802.3af-2003, Class 1 |
| Operation time of the battery | The batteries of RMS devices typically last 3 years under the optimal conditions: - Temperature = 20°C±5°C. - Wireless signal strength >-60 dBm. - Limited “read out data” events in the audit trail. |
| Log memory | 44000 of measurement for each channel |
| Startup time logger | < 20 s (RMS-Webservice (DHCP) < 40 s) |
| Measurement interval | 10 s ... 300 s |
| Adjustment | 1 or 2 point per channel |
| Sensor quality | - Detection of short circuit on the sensor connectors - Detection if no sensor connected |
| Temperature simulator function for validation | Fix values for each channel |
| Operational temperature logger | -40 ... 70 °C |
| Storage temperature | -40 ... 30 °C NOTE: Small storage temperature range is based on the battery life |
| Housing Material | PC-ABS |
| IP protection | IP65 |
| Weight | 240 g |