Modbus

Modbus TCP Server

Important:
Please read out the MODBUS byte order of the RMS device with RMS-CONFIG prior to any programming.

The device has a Modbus TCP server for direct connection to external devices. The data accessible via Modbus can be found in the table below.

Details
Protocol Modbus TCP
TCP Port 502

When reading out data from the input register select the index=-1.

Function 04 – Read Input Register

Register Parameter Data type
30001 serial number Unsigned 32 Bit
30002
30003 Probe serial number Always 0x00000000 Unsigned 32 Bit
30004
Register Parameter Data type
31001 Measurement value 1 Temperature value RTD 1 Float 32 Bit
31002
31003 Measurement value 2 Temperature value RTD 2 Float 32 Bit
31004

The temperature values are supplied only as float.

Float Swap Mode

Swap Mode is used to set how a Float32 is to be mapped to the Modbus data structure:

45   = 0x42340000 1000 = 0x447A0000

The RMS LAN devices (except the RMS-Converter) are based on a Little Endian memory organisation (The device firmware provides the MODBUS information starting with the most significant byte.)

31001 0 0x0000 31001 13378 0x3442
31002 13378 0x3442 31002 0 0x0000
31003 0 0x0000 31003 31300 0x7A44
31004 31300 0x7A44 31004 0 0x0000
Byte Swap (Mixed Endian) Byte and Word Swap (Big Endian)
31001 0 0x0000 31001 16948 0x4234
31002 16948 0x4234 31002 0 0x0000
31003 0 0x0000 31003 17530 0x447A
31004 17530 0x447A 31004 0 0x0000

Connection timeout: When TCP connection is open and for more than 30 seconds no communication is performed, the device automatically closes the TCP connection. In case of a communication interruption, this prevents the socket being blocked.

Additional MODBUS details can be found here: MODBUS in RMS.

Updated: 13.12.2023.