Device description

General information

Functions Description
Measured parameters Relative humidity and temperature using the ROTRONIC HYGROMER® HT-1 and a Pt100 temperature sensor
Humidity measurement range 0…100 %rh
Temperature measurement range Electronic: -40…+85 °C
Sensor head: -100…+200 °C1
Continuous Load: 190 °C max
Accuracy ±0.8 %rh, ±0.1 K at 10…30 °C
Diameter 15 mm (0.6")
Length Probe head: 144 mm
Cable length: 2 m (HCD-IC102)
Electronics and connector: 111 mm
Housing material PPS (black), stainless steel 1.4301
Standard filter non
Weight 450 g
Power Supply The HCD-IC requires a power supply of 3.3…15 VDC
Digital interface The HCD-IC probe has a UART interface with an logical level of 3.3 V
Modbus Measurement values and serial number can be read out if the probe is to be used out of the RMS - please see Modbus
Adjustment - 1-point or multi-point humidity adjustment/calibration
- 1-point or 2-point temperature adjustment/calibration
- Output the last adjustment/calibration, with date
- Produce an adjustment/calibration log
Sensor alarm The probe outputs a digital alarm if the sensor has a short circuit or open circuit. This function cannot be deactivated
Cable length maximum 3 m

1The peak load at 200 °C is 100 h maximum. Continuous Load 190 °C maximum.

Factory defaults

Configurable settings Factory default
Unit of measurement (metric/English) Metric
Psychrometric calculation None
Communication protocol MODBUS
MODBUS address 0
Device name HCD-IC102
Functions Factory default
Humidity/temperature adjustment Factory adjusted
Write-protection for device settings Disabled
Simulator mode Disabled