| ID | Description |
|---|---|
| RMS-2259 | New software architecture. Details: the RMS software structure has been improved. The main benefits are based upon the scaling for the cloud: - The system is improved for larger amounts of devices and measuring points. - Stability improvement. |
**Improvements **
| ID | Description |
|---|---|
| RMS-2533 | Change gateway concept: - The average of the last 5 RSSI values is taken. - If one value is missing, the average RSSI value will be reduced by 10. - The gateway will be switched if the difference is >=15dB (or 10dBm if the RSSI <-90dBm). |
| RMS-2532 | Adapted the download commands: - The download commands will be sent over a period of 3 days. After 3 days a data gap will be created. - The download command will be sent at increasing intervals. The intervals will be doubled until a maximum of 2 hours (ex: 1, 2, 4, 8, 16, 32, 64, 120, 120, 120...). - Within 3 days, the around 40 download trials will happen. |
| RMS-2534 | The RMS-Monitor has been removed from the system. |
| RMS-2586 | The RMS software will work with the latest Google Chrome update (Javascript). |
Bug Fixes¶
| ID | Description |
|---|---|
| RMS-2389 | Typo: "AES Key Refresh" not "AES Key Refeseh". |
| RMS-2383 | Unit missing in PDF reports. |
| RMS-2294 | Property definitions cannot be overwritten. |
| RMS-2293 | Firmware update: "NoPropertyData". |
| RMS-2234 | Not possible to archive a measuring point. |
| RMS-1178 | MPT-number not available with an SMS. |
| RMS-1117 | Data from an unassigned gateway will not be taken (if encrypted). |
| RMS-1093 | Archiving crashed when copying measuring data. |
| RMS-1051 | Software exception: UpdateCalculations. |