V2.5.0.23166

Phase 3 release date: 13.10.2023.

New Features

ID Description
RMS-543A Calibration overview page under the calibration menu. This change also covers the requirements from RMS-3093.
RMS-546A A batch release report can be created under data analysis. This change also covers the requirements from RMS-1089.
RMS-654A The measuring point colour for reports can be defined under the measuring point settings. This change also covers the requirements from RMS-3136/2251/960/808.
RMS-671A Login via SSO (SAML2.0). This change also covers the requirements from RMS-2979/2892.
RMS-893A Two step sign off for all user reports. This change also covers the requirements from RMS-3079/2788/1631.

Improvements

ID Description
RMS-499A Update event report template to include all filters. Limited selection within the export function in the events page.
RMS-649A All PDF reports are in PDF/A format, ISO19005-1 requirement. This change also covers the requirements from RMS-3145/2796.
RMS-650A Update of all RMS report logos. This change also covers the requirements from RMS-2915.
RMS-651A All report names can be extended with a timestamp. This change also covers the requirements from RMS-941.
RMS-652A User ID added to all report footers. This change also covers the requirements from RMS-2978.
RMS-653A Select the option to show separate charts, with various groupings in the report template. This change also covers the requirements from RMS-3128/3100/2887/2327.
RMS-656A Define multiple time periodes for the daily min/max table in the report template. This change also covers the requirements from RMS-3135.
RMS-657A Select the option to see the alarm limits, even when they are not active in the report template. This change also covers the requirements from RMS-3129.
RMS-658A Select the option to see related events in the report template. This change also covers the requirements from RMS-2680/2661/2480/2362/1188.
RMS-659A The seconds are included in the time stamps in reports. This change also covers the requirements from RMS-2390.
RMS-660A Select the option to see the calibration due date in the report template.
RMS-661A Select the option  to see the hysteresis and delay time in the report template. This change also covers the requirements from RMS-2691.
RMS-662A The confirmation details are shown in the event report.
RMS-663A Select the send time and day for the sending of an E-Mail report. This change also covers the requirements from RMS-3130/3080.
RMS-725A Calibration page and certificate updated:
1) Update input calibration to as found.
2) Update Output calibration to as left.
3) Separate the as found and as left values.
a) Summary As Found.
b) Summary As Left.
4) Add the calibration date to the summary.
5) Update Measurement to UUT (unit under test).
6) Update Difference to Error.
This change also covers the requirements from RMS-3158/2996/1171.
RMS-819A Add the calibration due date directly in the measuring point setup.
RMS-829A Under notifications, there is the possibility to select each different type of each event.
RMS-881A Chart and Table improvement for DI/DO.
RMS-884A Select input units for analog measuring points.
RMS-892A Calibrate multiple measuring points simultaneously. This change also covers the requirements from RMS-796.
RMS-906A LDAP server login with username and password.
RMS-1224A Mapping report improvement:
1. The lower and higher points are shown in blue (lowest) and red (highest).
2. The deviation show from/to.
3. Removal of the alarm level for the device characterisation and replaced with the setpoint and maximum error alarm.
4. When information is not completed, then --- is shown in the report.
5. Correction of the statistical overview in the device characterisation.
6. Calibration certificate and uncertainty saved per measuring point.
RMS-1264A Update report to ensure that the table header is included on each page.
RMS-1299A RMS saves all DO and DI measure values for every interval. When a timeout occurs, the missing data, unless a change happened will remain empty.
RMS-1379A The limit of 20 measuring points for converter devices has been removed.
RMS-1476A Select the option to add trends to graphics in the report template. This change also covers the requirements from RMS-3137.
RMS-1522A UserName added to the report footer. This change also covers the requirements from RMS-2978.

Bug Fixes

ID Description
RMS-1271A The title text overlaps into the rest of the screen.
RMS-1282A Within the mapping function, when selecting a month, the arrows are missing.
RMS-1288A When selecting the option to view the diagnose function, the menu and groups are not visible.
RMS-1289A Data analyse alarm limit when the system is setup in °F, Min/Max value wrong (low instead of high).
RMS-1292A Group name issues when using special characters (°, ö, ä, ü...). This change also covers the requirements from RMS-3054.
RMS-1294A RMS-MLOG-B is reported as RMS-LOG-868 in the device inventory. Read out device settings required prior to extracting the device inventory.
RMS-1297A Graph view of the data analyse for the DI, DO, virtual button and tristate button are shown in steps.
RMS-1302A The deviation report does not show related device events.
RMS-1387A User password reset not working when user was blocked by the system.
RMS-1388A Wrong alarm zones in mapping report chart.
RMS-1392A HTML of the event in included in the audit trail event report.
RMS-1407A Wrong alarm text in mapping report PDF.
RMS-1440A SMS notification of a database offline event is not documented in the audit trail.
RMS-1525A Datagap analysis does not work for analog measuring points.
RMS-1558A PDF mapping reports cannot be validated.
RMS-1565A CreatePdfReport: System.OverflowException.
RMS-1571A The call notification does not continue when an inhibition is cancelled.
RMS-1596A When a user has no rights to see the system setup page, a server load error appears instead of access denied.
RMS-1611A Access denied when looking at show device properties if the delete right is not selected.
RMS-1696A Labels not shown in layout setup.
RMS-1816A Adjust page fails without Dataview user right.
RMS-1845A Area report: exception when a measuring point doesn't have data for a certain time frame.
RMS-1866A Measuring point unit missing in the configuration report.
RMS-1874A Phone notification not stopping.
RMS-1878A Wrong conversion to particle/ft3.

Updated: 22.11.2023.