RMS-CONVERTER Open Issues¶
The RMS-CONVERTER open issues are only valid as of the most actual firmware and software versions.
| N/A | RMS-CONVERTER-100 and all Rotronic devices with time displayed. | The data logger time will always show the UTC time. | RMS system requirement. | No solution available. |
|---|---|---|---|---|
| RMS-CONVERTER-2594 | RMS-CONVERTER-100 and HygroLog and HL-NT products. | The HygroLog NT/HL-NT freezes. The user needs to remove all power supplies in order to restart the HygroLog NT/HL-NT. | HygroLog NT/HL-NT issue. | No solution available. |
| RMS-CONVERTER-2638 | RMS-CONVERTER-100 and HygroLog and HL-NT products. | The HygroLog NT/HL-NT measuring points are frequently going into sensor error and recovering. | Currently not defined. | No solution available. |
| RMS-CONVERTER-2427 | RMS-CONVERTER-100 and HygroLog and HL-NT products. | Peaks on RMS and not on the HygroLog NT/HL-NT. | Currently not defined. | No solution available. |
| RMS-CONVERTER-2355 | RMS-CONVERTER-100 and HF8 products. | The switching of the relay creates an event "Probe replaced" and shows "Serial number -> 0" | Currently not defined. | No solution available. |
| RMS-CONVERTER-2496 | RMS-CONVERTER-100 V1.4 and HygroLog and HL-NT products. | Data gaps appear when reading out digital inputs. | RMS-CONVERTER-100 and RMS software issue. | RMS-CONVERTER-V1.5 and RMS V2.2. |
| RMS-CONVERTER-2476 | RMS-CONVERTER-100 V1.4 and HygroLog and HL-NT products. | Device & Server time synchronization is missing within the audit trail. Data gaps occur. | RMS-CONVERTER-100 and RMS software issue. | RMS-CONVERTER-V1.5 and RMS V2.2. |